Advertisement
Science

How to Adjust and Use a Compound Microscope

A compound microscope is basically a device capable of magnifying microscopic elements that may be invisible to normal human eyes. Let’s understand how to adjust and use a compound microscope.

By Swagatam
Desk Science
Reading time 5 min read
Word count 985
Medical Science Lab testing
How to Adjust and Use a Compound Microscope
Advertisement
Quick Take

A compound microscope is basically a device capable of magnifying microscopic elements that may be invisible to normal human eyes. Let’s understand how to adjust and use a compound microscope.

On this page

A microscope can be considered one of the greatest inventions and contributions to medical science. A microscope is an optical instrument with sets of intricately devised lenses capable of magnifying anything that may be impossible to visualize through normal human eyes.

Specifically these devices are utilized in laboratories to examine various harmful parasites or microbes like bacteria and viruses, as well as other biological specimens, including all types of human tissue.

Advertisement

Just like any other instrument a microscope too may have its own limitation as far as magnifying is concerned and is termed as the resolving power or the resolution of a microscope. An ordinary microscope normally may be quite inefficient in providing comprehensive details regarding a microbe or an infected tissue due to its low resolution. To overcome this a compound microscope is usually preferred which may have a resolution or a magnifying capacity of more than 1000 times. A small in-built light source is normally incorporated inside them for illuminating the specimen under scan and hence they are also known as compound light microscopes.

Their resolving power is typically around 0.2 µm, the wavelength of visible light being just about the double of this figure. Images obtained from this type of microscope are called photomicrographs.

Advertisement

The following section discusses how to adjust and use a compound microscope.

Setting Up the Microscope

The following stages are normally employed to set up a compound light microscope:

Advertisement
  • Initially, the resolving nose piece is moved to suitably lock the lower objective lens into position.

  • Next, looking at the objective lens and the stage sideways, the coarse focus shaft is turned to make the stage move near the objective and as close as possible to the slide without making physical contact.

    Advertisement
  • The light source and the diaphragm are now adjusted until optimum illumination is reached.

  • After this the coarse focus is gently turned to move the stage downwards until the view shifts into a broader focus. The fine adjustment control is further optimized for a perfect focus.

    Advertisement
  • Finally, the specimen slide is moved until it appears at the center of the frame, again the illumination and the diaphragm is adjusted to improve clarity.

How a Compound Microscope is Used in Laboratories

The following couple of test methods (normally employed for the diagnosis of disease causing pathogens) and the acquired test results using a compound microscope simply proves its efficiency and indispensability in the medical field. The examples will also help you to understand how to prepare a piece of tissue of examination under a compound light microscope.

Advertisement

Let’s analyze the utility of a Compound Light Microscope (CLM) through the following illustrations:

Identifying a Yeast Colony

**

Yeast Cell, Image
Yeast Cell, Image
**Sometimes it becomes difficult to differentiate between a yeast colony and a bacterial colony
Bacteria Cells, Image
Bacteria Cells, Image
macroscopically, during such occasions a wet mount slide of the colony is used for inspection and the following procedures are conducted:

Advertisement

The microscopic slide is first sanitized using sterile water,

A very small quantity of the specimen colony is diluted in a drop of sterile water using a sterile bacteriological inoculating loop,

Advertisement

The prepared slide is covered with a glass coverslip,

The slide is then inserted under a CLM for examination. It will be important to note that for a wet mount specimen an oil immersion lens is never used.

Advertisement

Observation: The magnified image of the sample slide under the CLM normally yields the following facts:

Yeast cells are more easily distinguishable due to the major difference in their size and shape from the bacterial cells.

Advertisement

They are much larger - about 10 times more than the bacterial cells, moreover yeast cells may be found to be oval in shape unlike the bacterial cells which are mostly elongated.

The important identifiable feature among yeast cells are their budding phases which is never found in bacterial cells.

Advertisement

You may also find that the heat of the CLM light source causes the yeast cells to wiggle. This “wiggling” of the cells is technically termed as the Brownian motion. However this may happen only if the examinations are done for prolonged periods.

Diagnosing Vaginal Infections

**

Motile Trophozoites Under Compound Microscope, Image
Motile Trophozoites Under Compound Microscope, Image
**A drop of sterile saline is placed over a clean glass slide,

Advertisement

A sample of the vaginal discharge is acquired over a sterile swab,

The specimen material in the swab is gently brought in contact with the sterile saline and stirred to let them mix,

Advertisement

Almost immediately the preparation is covered with the glass coverslip so that it DOES NOT dry up,

The above wet mount is investigated under a CLM in a rather reduced illumination (Do not use oil immersion lens for wet mounts),

Observation: The CM magnifications may normally reveal the following possible causes of infection:

Presence of yeast cells, indicating colonies of yeast vaganities,

If motile trophozoites of T. vaginalies are seen, the presence of trichomoniasis may be presumed,

If an accumulation of clue cells (squamous epithelial cells occupied with colonies of cocobacilli) is witnessed, will indicate a bacterial vaginosis.

In the above case a Gram stain preparation of the vaginal fluid is recommended and its inspection under the CM may specify the following results:

Presence of some Gram-positive bacilli or Lactobacillus spp.

Presence of many Gram-negative bacilli or Bacteroids spp. Many curved Gram-negative bacilli or Mobiluncus spp.

Presence of Gram-variable coccobacilli or Gardnerella vaginalis.

Identification of Anthrax Bacteria

Anthrax Bacteria, Image Under Compound Microscope
Anthrax Bacteria, Image Under Compound Microscope
A CLM is also used to identify Anthrax bacteria. Here, as explained above a drop of sample blood of the victim is placed and Gram stained over a glass slide. The specimen is then visualized under a CLM. If it’s anthrax, the magnified view instantly confirms this due to the presence of relatively larger sized bacteria. The presence of violet-stained rod shaped bacteria and polymorphal white blood cells along with a very few red blood cells further back-up an anthrax foothold.

References

https://books.google.co.in/books/

https://medent.usyd.edu.au/fact/scabies.html#laborat

https://www.precisionliftinc.com/reconn-anthrax.shtml

Keep Exploring

More from Science

How Reading Rewires Your Brain

How Reading Rewires Your Brain

Confucius said, “Without knowing the force of words, it is impossible to know men.” The largest part of your brain is …

Second-Hand Smoke Issues

Second-Hand Smoke Issues

What is Secondhand Smoke? Secondhand smoke consists of the plume of chemicals and burning agents that come off the tip …

Filed under
Medical Science
More topics
Lab testing
Advertisement